產品分類品牌分類
-
超聲波探傷標準試塊 著色滲透探傷劑/顯像劑/滲透劑/清洗劑 磁粉檢測試塊/磁粉探傷試塊 磁粉探傷耗材 移動式磁粉探傷儀 多用磁粉探傷儀 磁粉磁軛探傷儀/磁探儀 X射線報警儀/個人計量儀 暗室紅燈/三色燈/警告燈 超聲波探傷儀/測厚儀 退磁機/線圈退磁器 冷熱觀片燈/工業評片燈 強度計/特斯拉計/高斯計 數顯黑白密度計/密度片 超聲波探傷耗材 射線探傷耗材 磁粉、磁膏、磁懸液 高強度紫外線燈 熒光探傷燈 紫外線黑光燈/LED黑光燈/熒光探傷燈 工業X光膠片 JB行業標準試塊 GB國家標準試塊 CSK-IA超聲波試塊 焊縫自然缺陷試塊 超聲波檢測試塊 *標準試塊與其他*標準試塊 工業恒溫洗片機、干片機 進口 超聲波耦合劑 其它
PH復合電極中英文對照說明書
Model E-201-9 Combination Electrode Operation Instruction
E-201-9型復合電極使用說明
I. Application用途
Model E-201-9 electrode is combined by glass electrode and reference electrode for measuring Ph value in water. It is widely used in chemical light industry, pharmaceutical industry, dye-works and scientific institute where pH value needs measuring.
本電極是玻璃電極和參比電極組合在一起的復合電極,用于測量水溶液的pH值。它廣泛應用于化輕工業、制藥工業、染色工業和科研機構中需要檢測水溶液PH值的場合。
II. Specification 規格
1. Measuring range測量范圍: 0-14 pH
2. Measuring temperature測量溫度: 0-60℃
3. Zero potential零電位: 7±0.5pH(25℃)
4. Slope斜率: (PTS)≥98.5%(25℃)
5. Internal resistance內電阻:≤250MΩ(25℃)
6. Alkali error堿誤差 : 0.2pH(1M Na+ pH14) (25℃)
7. Response time響應時間: towards 95% equilibrium value, less than 1 sec./對于 95%均衡值,少于1秒。
III. Operation and maintenance使用和維護
1. Before measuring, electrode must be calibrated with standard buffer solution.
測量前,電極必須用標準緩沖液校準。
2. Sensitive glass bubble is not allowed to touch with hard substances. Electrode wills no effective if any rubbing is made.
敏感玻璃球泡不要與硬物接觸,任何破損和擦毛都會使電極失效。
3. When measurement finished, put electrode in the cap, this has a bit of KCL solution in it.
完成測量后,應將電極保護帽套上,帽內應放少量氯化鉀溶液。
4. Outer reference solution of combination electrode is 3 mol/L KCL solutions (Mix 20ml distilled water with 3 mol/L KCL). Solution can be poured from small hole under the cap.
復合電極的外參比補充液為3 mol/1 KCL (20ml蒸餾水和3 mol/L KCL混合),補充液可從上端小孔加入。
5. Plug of electrode must be kept clean and dry. Short circuit of two ends of the plug is prohibited.
電極的插頭必須保持清潔和干燥,防止輸出端兩端短路。
6. Electrode should be matched with pH meter that has high input impedance(≥1012Ω)
電極應與高輸入阻抗(≥1012Ω)的pH計配套。
7. Electrode mustn’t be dipped in distilled water, protein solution or acid fluoride solution for long time. It mustn’t’ touch with organic silicon grease.
電極避免長期浸在蒸餾水、蛋白質、酸性氟化物溶液中,并防止和有機硅油脂接觸。
8. If slope drops, dip sensitive bubble of the electrode in 4% HF solution for 3-5 sec. clean with distilled water, dip in KCL solution to make it renew.
如果斜率下降,把電極敏感球泡端浸泡在4%HF(氫氟酸)中3-5秒,用蒸餾水洗凈,浸泡在KCL溶液中使之復新。
9. If bubble is contaminated or liquid junction is blocked, sensitivity of electrode will be influenced. Clean with proper detergent.
被球泡受污染或液接界堵塞,電極靈敏度會受影響,此時應用適當清潔劑清洗。
10. Flowing table is reference for detergent selection:
清潔劑選擇看下表,供參考
Contaminated Substance污染物 | Detergent清洗劑 |
Inorganic metal oxide有機金屬氧化物 | Less than 1 mol/1 acid 低于/M鹽酸 |
Organic grease substance有機脂類物質 | Weak alkalinity稀皂液或洗滌劑 |
Resin substance樹脂高分子物質 | Alcohol, acetone, ether |
Protein blood deposit蛋白質血球沉淀物 | Acidity enzyme solution酸性酶溶液 |
Dyestuff substance染料類物質 | H2O2 diluted bleaching powder 過氧化氫,稀漂白液 |
Effective storage time:12 months